| Vacuum Chamber PM Technique | ||
| LAM 2300 Exelan Flex Oxide Etch Chamber | ||
| OBJECTIVE: | ||
| TO EFFECTIVELY PM THE LAM 2300 EXELAN FLEX OXIDE ETCH CHAMBER IN A TIMELY MANNER, WHILE IMPROVING TOOL RECOVERY AND PARTICLE PERFORMANCE | ||
| Vacuum Chamber: | LAM 2300 EXELAN FLEX OXIDE ETCH | |
| Vacuum Chamber Process Residue: | PROCESS INDUCED RESIDUE | |
| Vacuum Chamber Components: | CHAMBER, CHAMBER LINERS AND ASSOCIATED PARTS |
| Old Procedure: | 2+ hours using silicon carbide scrubbing pads with ethanol (or equivalent final wipe solvent) 100+ wipes |
|
| Tool recovery: 6 to 8 hours conditioning wafers | ||
| New Procedure: | 1 hour using 800D ScrubPAD, DI water & ethanol (or equivalent final wipe solvent) | |
| Tool recovery: <4 Hours conditioning wafers | ||
![]() |
||||
| Vacuum Chamber Products: | ||||
| LAM2300 EXELAN OXIDE ETCH PM KIT | ||||
| PM Kit P/N: HT4500-LAM23-OX1 | ||||
| (1) HT4754 UltraSOLV® Sponge | ||||
| (2) HT4580D-10-1 800 Grit Diamond ScrubPAD | ||||
| (4) HT1511FC-5 MiraSWABS® (20pc) | ||||
| (1) HT5790S-25 MiraWIPES® (25pc) | ||||
| LAM 2300 Exelan Flex Oxide Etch PM Procedure: | ||||||
| View “How to” instructional videos on http://www.foamtecintlwcc.com/video/ | ||||||
| Step 1: | Using proper procedures and safety guidelines, shutdown and prepare LAM 2300 Exelan Flex Chamber for wet clean | |||||
| Step 2: | Properly stage a container of DI water next to the chamber and place a Foamtec International HT4580D ScrubPAD and HT4754 UltraSOLV® Sponge into the container (See Fig 1) | |||||
![]() |
||||||||
Fig 1: ScrubPAD and UltraSOLV® Sponge in container of DI water |
||||||||
| NOTE: | MAY SUBSTITUTE A SMALL AMOUNT OF DI WATER TO MOISTEN UltraSOLV® SPONGE AND ScrubPAD AS NECESSARY (See Fig 2 & 3) | |||||||
|
||||||||
Fig 2: UltraSOLV® Sponge, ScrubPAD and small amount of DI water |
Fig 3: Dampening UltraSOLV® Sponge with DI water
|
|||
| Step 3: | Use the lightly dampened UltraSOLV® Sponge and wipe Oxide ETCH Chamber liners (See Fig 4) |
![]() |
|||
Fig 4: Dampened UltraSOLV® Sponge wiping chamber liners
|
|||
| Step 4: | Use a lightly dampened 800 Grit Diamond ScrubPAD and scrub off deposition from Oxide ETCH Chamber liner. It is important to keep area a little moist with DI water (See Fig 5 & 6) | ||
![]() |
![]() |
|||
Fig 5 & 6: Lightly dampened 800 Grit Diamond ScrubPAD scrubbing chamber liner; keep slightly moist with DI water |
||||
| Step 5: | Before oxide deposition has a chance to dry back onto the chamber liner, use lightly dampened UltraSOLV® Sponge and wipe deposition from chamber liner (See Fig 7 & 8) | ||||
![]() |
![]() |
|||
Fig 7: Lightly dampened UltraSOLV® Sponge preparing to wipe chamber liner
|
Fig 8: Clean chamber liner after UltraSOLV® Sponge |
|||
| Step 6: | As Diamond ScrubPAD appears to load up with deposition, pull ScrubPAD across damp UltraSOLV Sponge. This will help keep ScrubPAD effectively removing oxide deposition from chamber liner (See Fig 9, 10 & 11) | ||
![]() |
![]() |
![]() |
||||||||||
Fig 9: ScrubPAD loaded with deposition |
Fig 10: Pull ScrubPAD across UltraSOLV® Sponge |
Fig 11: Unloaded ScrubPAD |
||||||||||
| Step 7: | Continue to rinse UltraSOLV® Sponge in container of DI water as necessary to keep UltraSOLV® Sponge slightly moist and free of deposition (See Fig 12 & 13) | ||
![]() |
![]() |
|||
Fig 12: UltraSOLV® Sponge loaded with deposition
|
Fig 13: UltraSOLV® Sponge free of deposition after rinse in DI water
|
|||
| Step 8: | Continue to repeat this SCRUB – WIPE – RINSE procedure outlined in steps 4 through 7 for the remainder of the LAM 2300 Oxide ETCH parts (See Fig 14, 15 & 16) | ||
![]() |
![]() |
![]() |
||||||||||
Fig 14, 15 & 16: Continuing SCRUB – WIPE – RINSE procedure on remaining Oxide ETCH Chamber parts |
||
| Step 9: | Using same technique described above, take 800 Grit Diamond ScrubPAD and gently remove dark build-up on heater coils (See Fig 17) | ||
![]() |
|||
Fig 17: 800 Grit Diamond ScrubPAD removing build-up on heater coils |
|||
| FINAL WIPE PROCEDURE: | ||
IMPORTANT NOTE |
||
IN ORDER TO HELP WITH AN EFFECTIVE TOOL RECOVERY, FOAMTEC INTERNATIONAL WIPE DOWN PROCEDURE MUST BE FOLLOWED WITH ALL HT5790S MiraWIPES® AND HT1511FC MiraSWABS®. THE MICRO-FIBER CHARACTERISTICS OF THESE PRODUCTS HELP REMOVE MORE DEPOSITION FROM THE PARTS THAN ANY OTHER STANDARD FAB WIPER |
||
Figure below shows how much more deposition the Foamtec International MiraWIPE® can remove from a critical surface compared to the standard fab wiper, making the MiraWIPE® FINAL WIPE PROCEDURE the most CRITICAL STEP of the PM procedure (See Fig 18a & 18b) |
||
![]() |
||
Fig 18a: Current fab wiper after completely wiping the LAM 2300 |
Fig 18b: Particles picked up using HT5790S MiraWIPES® after completely wiping with current fab wiper |
|
![]() |
![]() |
|
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY |
||
| Step 10: | After scrubbing heater coils use the HT1511FC MiraSWAB® dampened with alcohol and clear out all openings on heater plate (See Fig 19 & 20) |
![]() |
![]() |
|||
Fig 19: HT1511FC MiraSWAB® clearing out openings on heater plate
|
Fig 20: Additional deposition MiraSWAB® was able to remove from openings
|
|||
| Step 11: | While using the HT1511FC MiraSWABS® to clear out all the openings within the heater coils, also use the HT5790S MiraWIPES® dampened with alcohol and proceed to wipe the entire heater plate, concentrating on the tight corners throughout the heater plate (See Fig 21 & 22) | |||
![]() |
||||
Fig 21: Wiping heater plate with HT5790S MiraWIPE® and alcohol |
||||
![]() |
|||
Fig 22: Clean heater plate coils after gentle scrub and wipe
|
|||
| Step 12: | Following the same procedure with the HT1511FC MiraSWABS® and HT5790S MiraWIPES® with alcohol, perform an entire wipe of the remaining Oxide ETCH parts prior to placing back into the Oxide ETCH Chamber (See Fig 23) | ||
![]() |
|||
Fig 23: HT5790S MiraWIPE® performing final wipe on all chamber parts |
|||
| NOTE: | THE FOLLOWING SHOWS HOW MUCH MORE DEPOSITION THE MiraWIPE® WAS ABLE TO REMOVE FROM A SMALL SECTION OF THE SILICONE RING AFTER A COMPLETE WIPED DOWN WITH THE STANDARD FAB WIPER (See Fig 24 & 25) | ||
![]() |
![]() |
||||||||||
Fig 24 & 25: THE IMPORTANCE OF USING THE HT5790S MiraWIPES®
|
|||||||||||
| CHAMBER CLEAN: | |||||||||||
| Step 13: | Take the additional 800 Grit Diamond ScrubPAD and using the same technique described above proceed to perform a chamber scrub on the LAM 2300 Exelan Flex Oxide Chamber |
| NOTE: | Ensure to concentrate on the heavy deposition areas that are not covered by the chamber liners, such as the sides of the e-chuck (See Fig 26) |
![]() | |||
Fig 26: 800 Grit Diamond ScrubPAD gently scrubbing edge of e-chuck | |||
| Step 14: | When entire chamber scrub is complete, use the same technique described above and use remaining HT1511FC MiraSWABS® and HT5790S MiraWIPES® with alcohol and perform a complete chamber wipe down (See Fig 27 & 28) |
![]() |
![]() |
|||
Fig 27: Applying alcohol to HT5790S MiraWIPE®
|
Fig 28: Performing final wipe of LAM Oxide Chamber
|
|||
| TOOL RECOVERY: | ||||
| Step 15: | Follow proper tool recovery guidelines as outlined by LAM Research Corporation | |||