LAM 490 Auto Etch Chamber
 
OBJECTIVE:
TO EFFECTIVELY PM THE LAM 490 CHAMBER IN A TIMELY MANNER, WHILE IMPROVING TOOL PERFORMANCE AND REDUCING PARTICLE DEFECTS
 
Vacuum Chamber: LAM 490 AUTO ETCH
Vacuum Chamber Process Residue: PROCESS INDUCED RESIDUE
Vacuum Chamber Components: PROCESS CHAMBER AND CERAMIC RING
 
Old Procedure: > 4 hours with Scotch-Brite™ & DI water
PROBLEMS: Not able to remove by-product in a timely manner and particle issues due to the use of Scotch-Brite™
New Procedure: 2 hours using 280 Grit Diamond ScrubPAD, UltraSOLV® Sponge, MiraSWABS® & MiraWIPES®
BENEFITS: ABLE TO REMOVE MORE BY-PRODUCTS AND NOT LEAVE BEHIND SCOTH-BRITEā„¢ PARTICLES HELPING REDUCE PARTICLE DEFECTS AND IMPROVE TOOL RECOVERY
 
Foamtec Products
Vacuum Chamber Products:
PM Kit P/N: HT4500-LAM490
(1) HT4528D-10-1 280 Grit Diamond ScrubPAD
(1) HT4754 UltraSOLV® Sponge
(2) HT5790S-5 MiraWIPES® (10 MiraWIPES®)
 
Not included in PM Kit
  (1) HT1511FC-5 MiraSWABS® (5 MiraSWABS®)
(1) HT179028D-1 280 Grit ScrubTIP® Swab
 
 
LAM 490 Chamber PM Procedure:
View “How to” instructional videos on http://www.foamtecintlwcc.com/video/
 
Step 1: Using proper procedures and safety guidelines, remove associated parts from LAM ETCH chamber
 
Step 2: Stage a container of DI water next to the process tool and place the HT4528D 280 Grit Diamond ScrubPAD and HT4754 UltraSOLV® Sponge into the container (See Fig 1)
 
 
Place Sponge and ScrubPAD in DI water
   
Fig 1: Placing UltraSOLV® Sponge and 280 Grit Diamond ScrubPAD into DI water
MiraSWABS
 
Step 3: Place clean room wipes dampened with DI water in the chamber. Let chamber soak while taking shower head off of chamber lid, approx. 15 minutes (See Fig 2)
 
Fab wipes soaked in DI water
   
Fig 2: Standard fab wipes soaked with DI water
 
 
Step 4: Remove shower head and ceramic ring from chamber lid. Soak ceramic ring in DI water (See Fig 3)
 
Remove shower head and ceramic ring
Fig 3: Remove shower head and ceramic ring
 
Step 5: Take the UltraSOLV® Sponge from the DI water and wring out most of the water. Wipe the chamber down to start removing the process residue (See Fig 4)
 
Wipe chamber with UltraSOLV Sponge
 
Fig 4: Wipe chamber with UltraSOLV® Sponge
 
Step 6: Take lightly dampened 280 Grit Diamond ScrubPAD and begin to scrub the process residue from the chamber (See Fig 5)
 
Scrubbing chamber  
Fig 5: Start scrubbing process build up from the process chamber
 
Step 7: As the HT4528D 280 Grit Diamond ScrubPAD begins to load up with deposition, pull ScrubPAD across dampened UltraSOLV® Sponge to properly unload ScrubPAD (See Fig 6, 7 & 8)
 
ScrubPAD loaded with deposition
Pull ScrubPAD across Sponge
Unloaded ScrubPAD
Fig 6: ScrubPAD loaded with deposition
Fig 7: Pull ScrubPAD across UltraSOLV® Sponge
Fig 8: Unloaded ScrubPAD
 
Step 8: For the corners and areas with heavy build up use the HT179028D 280 Grit ScrubTIP® Swab lightly dampened with DI water (See Fig 9)
 
ScrubTIP used to scrub build-up
 
Fig 9: Diamond ScrubTIP® used to scrub heavy build up
 
Step 9: Wipe the chamber periodically with the UltraSOLV® Sponge during the scrub process. When the sponge becomes loaded up with process residue wring it out in the DI water (See Fig 10 & 11)
 
Loaded UltraSOLV Sponge
UltraSOLV after rinse
Fig 10: Loaded-up UltraSOLV® Sponge
Fig 11: UltraSOLV® Sponge AFTER rinse
 
Step 10: Follow steps 5 through 9 to continue scrubbing all parts of the chamber until all process residue has been removed. Ensure port areas are cleaned well (See Fig 12 & 13)
Clean chamber port area
 
Cleaned chamber port areas
Fig 12 & 13: Ensure chamber port areas are cleaned
Before After
 
Step 11: Remove ceramic ring from the DI water it has been soaking in and scrub the remaining process residue off of it using the HT4528D 280 Grit Diamond ScrubPAD
 
FINAL WIPE PROCEDURE:
 
IMPORTANT NOTE
 
THE USE OF HT5790S MiraWIPES® AND HT1511FC MiraSWABS® DURING FINAL WIPE PORTION OF PROCEDURE IS A CRITICAL STEP TO EFFECTIVELY REMOVING PARTICLE DEFECTS FROM THE LAM ETCH CHAMBER
 
Figure below shows how much more deposition the Foamtec International MiraWIPE® can remove from a critical surface compared to the standard fab wiper, making the MiraWIPE® FINAL WIPE PROCEDURE the most CRITICAL STEP of the PM procedure (See Fig 14a & 14b)
 
MiraWIPES are more effective
Fig 14a: Current fab wiper after completely wiping the chamber
Fig 14b: Particles picked up using HT5790S MiraWIPES® after completely wiping with current fab wiper
Red Arrow
Red Arrow
 
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY
 
Step 12: Fold the HT5790S MiraWIPE® into quarters and dampen with IPA
 
Step 13: With the IPA dampened MiraWIPE® wipe down all areas of the chamber, ensuring to refold the MiraWIPE® as necessary to expose a clean side of the MiraWIPE® as you are wiping the parts (See Fig 15)
 
Use MiraWIPE to wipe chamber
Fig 15: Use the MiraWIPE® Wiper to wipe out the process chamber
 
Step 14: After initial wipe down of the process chamber use the HT1511FC MiraSWABS®, dampened with IPA, to clean out holes in the pump plate. Ensure all holes are cleaned out (See Fig 16 & 17)
 
MiraSWABS cleaning holes in pump plate  
Fig 16: Use the MiraSWABS® to clean residue from holes in pump plate
 
  MiraSWABS pick up more residue
Fig 17: Amount of residue left behind from the use of Scotch-Brite™ in the past the MiraSWABS® are able to clean from holes in pump plate
 
Step 15: Perform final wipe down of the process chamber and any part installed back into the Chamber using MiraWIPE® dampened with IPA
 
NOTE: Replace with a new dampened MiraWIPE® as necessary
 
Step 16: Follow steps to reinstall chamber parts and bring process tool back into production