Vacuum Chamber PM Technique
LAM 9600 Metal ETCH WET-Strip Procedure
OBJECTIVE:

TO PM THE LAM 9600 METAL ETCH CHAMBER IN AN EFFECTIVE AND TIMELY MANNER WHILE IMPROVING PARTICLE PERFORMANCE, TOOL RECOVERY AND MAXIMIZING TOOL UPTIME

 
Vacuum Chamber: LAM METAL ETCHER
Vacuum Chamber Process Residue: PROCESS INDUCED RESIDUE
Vacuum Chamber Components:

METAL ETCH CHAMBER, VAT VALVE, CHAMBER LID AND CHAMBER PARTS

Solvent:

DI WATER, IPA (ONLY)

Vacuum Chamber Products:
  • (1) HT4528D-10  280 Grit Diamond UltraSOLV® ScrubPAD
  • (1) HT4580D-10  800 Grit Diamond UltraSOLV® ScrubPAD
  • (2) HT111328D-1  280 Grit Diamond UltraSOLV® Straight ScrubTIP®
  • (2) HT111380D-1  800 Grit Diamond UltraSOLV® Straight ScrubTIP®
  • (2) HT174913PD-1  1350 Grit Diamond 3” ScrubTIP®
  • (1) HT4754 UltraSOLV® Sponge
  • (2) HT1790-5 UltraSOLV® Swab
  • (2) HT1750-5 UltraSOLV® Swab
  • (2) HT5790S-25 MiraWIPE® Wipers
LAM 9600 Metal Etch Wet-Strip PM Procedure:
View “How to” instructional videos on http://www.foamtecintlwcc.com/video/
Step 1: Using proper procedures and safety guidelines properly shutdown and vent entire Metal ETCH chamber
Step 2:

Reduce ETCH chamber heater settings to 35oC to 40oC

 
Step 3:

Remove slit valve doors and place in container with DI water to soak (See Fig 1a )

      Slit valve door & Foamtec products & DI water      
   

Fig 1a: Slit valve door

Fig 1b: Foamtec products: Diamond ScrubPads, Sponge & DI water

   
Step 4: Thoroughly moisten all areas within chamber with DI water using HT4754 UltraSOLV® Sponge; including slit valves, chamber walls, vat valve and chamber lid
Step 5: Allow Metal ETCH chamber to soak with DI water for one hour
Step 6: Place HT4754 UltraSOLV® Sponge, UltraSOLV® 280 Grit Diamond ScrubPAD and 280 Grit Diamond ScrubTIP® in container with approximately 1 liter of DI water (See Fig 1b)
Step 7: Soak and ring out the UltraSOLV® Sponge and proceed to wipe down all areas throughout Metal ETCH chamber (this will remove flakes and gross buildup).  Continue to repeat process until UltraSOLV® Sponge will no longer remove deposition (See Fig 2)
Sponge
Fig 2 : Pre-wipe chamber with UltraSOLV® Sponge
Step 8: With UltraSOLV® Sponge, dampen and moisten a 6” to 8” scrubbing area within the chamber
   
Step 9: With dampened UltraSOLV® 280 Grit Diamond ScrubPAD, scrub off deposition from moistened area (See Fig 3)
 
Scrubbing deposition
Fig 3 : Scrub off deposition using Diamond ScrubPAD
Step 10: As deposition begins to release from chamber areas, remove by taking the dampened UltraSOLV® Sponge and wiping the affected area (See Fig 4)
 
Wiping area with Sponge
Fig 4: Wiping the scrub areas using the UltraSOLV® Sponge
Step 11: Continue to repeat process throughout entire chamber until all deposition is removed (see notes for Special Cleaning Tasks)
Step 12: As UltraSOLV® ScrubPAD begins to load-up with deposition; it can be cleaned off by dragging across UltraSOLV® Sponge in one direction (See Fig 5, 6 & 7)
 
Loaded ScrubPAD
Pull ScrubPAD across Sponge
Unloaded ScrubPAD
Fig 5: ScrubPAD loaded with deposition
Fig 6: Pull ScrubPAD across UltraSOLV® Sponge
Fig 7: Unloaded ScrubPAD
Step 13: As UltraSOLV® Sponge begins to load-up with deposition; it can be rinsed off and rung-out in container of DI water (See Fig 8)
Unloading Sponge
Fig 8: Unloading the UltraSOLV® Sponge with DI water
Step 14: As water becomes filled with deposition, recommend disposing dirty DI water in appropriate hazardous waste collection tank and replacing with fresh DI water
 
Step 15: After all areas within entire chamber have been effectively cleaned, take UltraSOLV® 800 Grit Diamond ScrubPAD, and using the same technique, proceed to lightly scrub all areas scrubbed previously 
 
NOTE: Ensure to wipe chamber with a freshly rinsed out UltraSOLV® Sponge to prep the chamber for FINAL IPA WIPE-DOWN
 

SPECIAL TASKS (VAT VALVE, TIGHT CORNERS, O-RING GROOVES, MANOMETERS & REMOVABLE PARTS):

VAT VALVE:

 
Recommend cleaning VAT valve after chamber clean, as UltraSOLV® 280 Grit Diamond ScrubPAD will be softer, more flexible and easier to work with after cleaning ETCH chamber
 
Step 16: Take a DI water dampened HT4754 UltraSOLV® Sponge and proceed to reach into VAT valve from top and bottom to thoroughly wipe out the VAT valve (See Fig 9)
 
Wiping VAT valve
Fig 9 : Wipe-down VAT valve using UltraSOLV® Sponge
Step 17: Using modified extender tool, slide UltraSOLV® Sponge down into VAT valve to wipe areas not able to reach previously
 
Step 18: Using the dampened UltraSOLV® 280 Grit Diamond ScrubPAD and the modified extender tool, slide the ScrubPAD into the VAT valve to effectively clean down into the valve (See Fig 10)
Scrub VAT using extender
Fig 10 : Scrub down into the VAT valve using the Diamond ScrubPAD
NOTE: Use ScrubPAD after cleaning chamber as pad will soften & be easier to use in VAT valve
 
Step 19: Using same technique, wipe the inside of the VAT valve with the HT4754 UltraSOLV® Sponge
 
Step 20: Repeat sequence as necessary to remove as much deposition as possible
 
Step 21: In order to clean the tight corners and seals within the VAT valve still not reachable, moisten the 280 Grit Diamond ScrubTIP® and scrub the affected areas (See Fig 11)
 
Clean tight corners and edges using ScrubTIP
 
Fig 11:  Clean tight corners and edges using the Diamond ScrubTIP®
 
 
 
Step 22: Using the UltraSOLV® Foam Swabs in the same manner as the UltraSOLV® Sponge, wipe out the loosened deposition in the tight corners and seals within the VAT valve)
 
Step 23: Take UltraSOLV® 800 Grit Diamond ScrubPAD to perform final scrub and wipe out thoroughly using freshly rinsed UltraSOLV® Sponge
 
Step 24: With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout VAT valve
 
TIGHT CORNERS, EDGES & O’RING GROOVES:
 
Step 25: Dampen the 280 Grit Diamond and 800 Grit Diamond ScrubTIPS® with DI water and proceed to scrub all hard to reach tight corners and edges (See Fig 12 & 13)
 
     
 

Fig 12: 280 Grit Diamond ScrubTIP® cleaning track on bottom of Chamber

Fig 13: Track on bottom of Chamber, left side - Before ScrubTIP®;            right side - After ScrubTIP®

   
Step 26: In all areas the UltraSOLV® Sponge may not reach, use the UltraSOLV® Swabs in the same manner as the UltraSOLV® Sponge and wipe out the loosened deposition in the tight corners and edges
Step 27:

For o-ring grooves, use 1350 Grit Diamond ScrubTIPS® provided (See Fig 14)

   
ScrubTIP used to polish grooves
 
Fig 14:  Use 1350 Grit Diamond ScrubTIPS® to polish o-ring grooves
 
 
 
Step 28: Follow o-ring groove with similar size UltraSOLV® Swabs for wipe-down
 
Step 29: With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout tight corners, edges and o-ring grooves effectively as possible
 
MANOMETERS:
 
Step 30: Use dampened 280 Grit Diamond ScrubTIP® to reach into manometer plumbing and remove buildup within manometer ports (See Fig 15 & 16)
 
  Use ScrubTIP to clean all plumbing ports       Use ScrubTIP to clean all plumbing ports  
 
Fig 15 & 16: Use ScrubTIP® to clean all plumbing ports
 
Step 31: Use UltraSOLV® Swab to remove loosened deposition from within manometer (See Fig 17)
  Swabs cleaning manometer port      
 

Fig 17: Use UltraSOLV® Foam Swabs to wipe out deposition from manometer ports

 
Step 32: With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout manometers
REMOVABLE PARTS (SLIT VALVE DOORS, END PORT & END PORT WINDOW):
Step 33: On a designated work bench, wipe down parts with DI water dampened UltraSOLV® Sponge
Step 34: Use dampened UltraSOLV® 280 Grit Diamond ScrubPAD to remove deposition buildup on affected areas (See Fig 18 & 19)

  Use ScrubPAD to clean removable parts       Use ScrubPAD to clean removable parts  
 

Fig 18 & 19: Use ScrubPAD to clean removable parts

 
Step 35: Use associated ScrubTIP® to scrub tight corners, edges and o-ring grooves
Step 36: Use UltraSOLV® Sponge and UltraSOLV® Swabs to remove loosened deposition from associated areas (See Fig 20 & 21)

  Use UltraSOLV swabs for o-ring grooves       Use UltraSOLV Swabs for o-ring grooves  
 

Fig 20 & 21: Use UltraSOLV® Swabs for o-ring grooves

 
FINAL WIPE PROCEDURE:
IMPORTANT NOTE
This Portion of the Clean is CRITICAL for an Effective Tool Recovery.  As Final IPA Wipe Nears Completion, Use More and More Fresh MiraWIPES® to Reduce Chances of Introducing Particles Back Into the Chamber
 
Figure below shows how much more deposition the Foamtec International MiraWIPE® can remove from a critical surface compared to the standard fab wiper, making the MiraWIPE® FINAL WIPE PROCEDURE the most CRITICAL STEP of the PM procedure (See Fig 22a & 22b)
MiraWIPES are more effective
Fig 22a: Current fab wiper after completely wiping the LAM 9600
Fig 22b: Particles picked up using HT5790S MiraWIPES® after completely wiping with current fab wiper
Red Arrow
Red Arrow
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY
Step 37: With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout removable parts
 
Step 38: Using a single HT5790S MiraWIPE®, folded into quarters, moisten with IPA (See Fig. 23)
 
Final wipe-down
 
 
Fig 23:  Use MiraWIPE® with IPA for final tool wipe down
 
Step 39: Working from highest point down; proceed to effectively wipe out entire Metal ETCH chamber, refolding MiraWIPE® to expose fresh side of MiraWIPE®
 
Step 40: Continue to repeat this sequence until entire Metal ETCH chamber is effectively wiped out
TOOL RECOVERY:
Step 41: Ensure all PM procedures have been completed and checked off to the point of preparation for tool recovery
Step 42: Just prior to closing chamber lid, effectively wipe out entire chamber one last time using a fresh MiraWIPE® with IPA (See Fig 24)
Chamber
Fig 24: Wipe entire chamber with IPA & MiraWIPES® just prior to closing chamber lid
Step 43: Close and seal chamber lid