| Vacuum Chamber PM Technique | ||
| LAM 9600 Metal ETCH WET-Strip Procedure | ||
| OBJECTIVE: | ||
TO PM THE LAM 9600 METAL ETCH CHAMBER IN AN EFFECTIVE AND TIMELY MANNER WHILE IMPROVING PARTICLE PERFORMANCE, TOOL RECOVERY AND MAXIMIZING TOOL UPTIME |
||
| Vacuum Chamber: | LAM METAL ETCHER | |
| Vacuum Chamber Process Residue: | PROCESS INDUCED RESIDUE | |
| Vacuum Chamber Components: | METAL ETCH CHAMBER, VAT VALVE, CHAMBER LID AND CHAMBER PARTS |
| Solvent: | DI WATER, IPA (ONLY) |
|
| Vacuum Chamber Products: | |||
|
|||
| LAM 9600 Metal Etch Wet-Strip PM Procedure: | ||||||||||||||
| View “How to” instructional videos on http://www.foamtecintlwcc.com/video/ | ||||||||||||||
| Step 1: | Using proper procedures and safety guidelines properly shutdown and vent entire Metal ETCH chamber | |||||||||||||
| Step 2: | Reduce ETCH chamber heater settings to 35oC to 40oC |
|||||||||||||
| Step 3: | Remove slit valve doors and place in container with DI water to soak (See Fig 1a ) |
|||||||||||||
|
||||||||||||||
| Step 4: | Thoroughly moisten all areas within chamber with DI water using HT4754 UltraSOLV® Sponge; including slit valves, chamber walls, vat valve and chamber lid | |||||||||||||
| Step 5: | Allow Metal ETCH chamber to soak with DI water for one hour | |||||||||||||
| Step 6: | Place HT4754 UltraSOLV® Sponge, UltraSOLV® 280 Grit Diamond ScrubPAD and 280 Grit Diamond ScrubTIP® in container with approximately 1 liter of DI water (See Fig 1b) | |||||||||||||
| Step 7: | Soak and ring out the UltraSOLV® Sponge and proceed to wipe down all areas throughout Metal ETCH chamber (this will remove flakes and gross buildup). Continue to repeat process until UltraSOLV® Sponge will no longer remove deposition (See Fig 2) | |||||||||||||
![]() |
|||
Fig 2 : Pre-wipe chamber with UltraSOLV® Sponge |
|||
| Step 8: | With UltraSOLV® Sponge, dampen and moisten a 6” to 8” scrubbing area within the chamber | ||
| Step 9: | With dampened UltraSOLV® 280 Grit Diamond ScrubPAD, scrub off deposition from moistened area (See Fig 3) | ||
![]() |
|||
Fig 3 : Scrub off deposition using Diamond ScrubPAD |
|||
| Step 10: | As deposition begins to release from chamber areas, remove by taking the dampened UltraSOLV® Sponge and wiping the affected area (See Fig 4) | ||
![]() |
|||
Fig 4: Wiping the scrub areas using the UltraSOLV® Sponge |
|||
| Step 11: | Continue to repeat process throughout entire chamber until all deposition is removed (see notes for Special Cleaning Tasks) | ||
| Step 12: | As UltraSOLV® ScrubPAD begins to load-up with deposition; it can be cleaned off by dragging across UltraSOLV® Sponge in one direction (See Fig 5, 6 & 7) | ||
![]() |
![]() |
![]() |
||||||||||
Fig 5: ScrubPAD loaded with deposition |
Fig 6: Pull ScrubPAD across UltraSOLV® Sponge |
Fig 7: Unloaded ScrubPAD |
||||||||||
| Step 13: | As UltraSOLV® Sponge begins to load-up with deposition; it can be rinsed off and rung-out in container of DI water (See Fig 8) |
![]() |
|||
Fig 8: Unloading the UltraSOLV® Sponge with DI water
|
|||
| Step 14: | As water becomes filled with deposition, recommend disposing dirty DI water in appropriate hazardous waste collection tank and replacing with fresh DI water | ||
| Step 15: | After all areas within entire chamber have been effectively cleaned, take UltraSOLV® 800 Grit Diamond ScrubPAD, and using the same technique, proceed to lightly scrub all areas scrubbed previously | ||
| NOTE: | Ensure to wipe chamber with a freshly rinsed out UltraSOLV® Sponge to prep the chamber for FINAL IPA WIPE-DOWN | ||
SPECIAL TASKS (VAT VALVE, TIGHT CORNERS, O-RING GROOVES, MANOMETERS & REMOVABLE PARTS): VAT VALVE: |
|||
| Recommend cleaning VAT valve after chamber clean, as UltraSOLV® 280 Grit Diamond ScrubPAD will be softer, more flexible and easier to work with after cleaning ETCH chamber | |||
| Step 16: | Take a DI water dampened HT4754 UltraSOLV® Sponge and proceed to reach into VAT valve from top and bottom to thoroughly wipe out the VAT valve (See Fig 9) | ||
![]() |
|||
Fig 9 : Wipe-down VAT valve using UltraSOLV® Sponge |
|||
| Step 17: | Using modified extender tool, slide UltraSOLV® Sponge down into VAT valve to wipe areas not able to reach previously |
| Step 18: | Using the dampened UltraSOLV® 280 Grit Diamond ScrubPAD and the modified extender tool, slide the ScrubPAD into the VAT valve to effectively clean down into the valve (See Fig 10) |
![]() |
||||
Fig 10 : Scrub down into the VAT valve using the Diamond ScrubPAD |
||||
| NOTE: | Use ScrubPAD after cleaning chamber as pad will soften & be easier to use in VAT valve | ||||
| Step 19: | Using same technique, wipe the inside of the VAT valve with the HT4754 UltraSOLV® Sponge | ||||
| Step 20: | Repeat sequence as necessary to remove as much deposition as possible | ||||
| Step 21: | In order to clean the tight corners and seals within the VAT valve still not reachable, moisten the 280 Grit Diamond ScrubTIP® and scrub the affected areas (See Fig 11) | ||||
![]() |
||||||
Fig 11: Clean tight corners and edges using the Diamond ScrubTIP® |
||||||
| Step 22: | Using the UltraSOLV® Foam Swabs in the same manner as the UltraSOLV® Sponge, wipe out the loosened deposition in the tight corners and seals within the VAT valve) | |||||||||||
| Step 23: | Take UltraSOLV® 800 Grit Diamond ScrubPAD to perform final scrub and wipe out thoroughly using freshly rinsed UltraSOLV® Sponge | |||||||||||
| Step 24: | With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout VAT valve | |||||||||||
| TIGHT CORNERS, EDGES & O’RING GROOVES: | ||||||||||||
| Step 25: | Dampen the 280 Grit Diamond and 800 Grit Diamond ScrubTIPS® with DI water and proceed to scrub all hard to reach tight corners and edges (See Fig 12 & 13) | |||||||||||
|
||||||||||||
| Step 26: | In all areas the UltraSOLV® Sponge may not reach, use the UltraSOLV® Swabs in the same manner as the UltraSOLV® Sponge and wipe out the loosened deposition in the tight corners and edges | |||||||||||
| Step 27: | For o-ring grooves, use 1350 Grit Diamond ScrubTIPS® provided (See Fig 14) |
|||||||||||
![]() |
||||||
Fig 14: Use 1350 Grit Diamond ScrubTIPS® to polish o-ring grooves |
||||||
| Step 28: | Follow o-ring groove with similar size UltraSOLV® Swabs for wipe-down | ||||||||||
| Step 29: | With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout tight corners, edges and o-ring grooves effectively as possible | ||||||||||
| MANOMETERS: | |||||||||||
| Step 30: | Use dampened 280 Grit Diamond ScrubTIP® to reach into manometer plumbing and remove buildup within manometer ports (See Fig 15 & 16) | ||||||||||
|
|||||||||||
| Step 31: | Use UltraSOLV® Swab to remove loosened deposition from within manometer (See Fig 17) | ||||||||||||
|
|||||||||||||
| Step 32: | With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout manometers | ||||||||||||
| REMOVABLE PARTS (SLIT VALVE DOORS, END PORT & END PORT WINDOW): | |||||||||||||
| Step 33: | On a designated work bench, wipe down parts with DI water dampened UltraSOLV® Sponge | ||||||||||||
| Step 34: | Use dampened UltraSOLV® 280 Grit Diamond ScrubPAD to remove deposition buildup on affected areas (See Fig 18 & 19) | ||||||||||||
|
|||||||||||||
| Step 35: | Use associated ScrubTIP® to scrub tight corners, edges and o-ring grooves | ||||||||||||
| Step 36: | Use UltraSOLV® Sponge and UltraSOLV® Swabs to remove loosened deposition from associated areas (See Fig 20 & 21) | ||||||||||||
|
|||||||||||||
| FINAL WIPE PROCEDURE: | ||
IMPORTANT NOTE |
||
| This Portion of the Clean is CRITICAL for an Effective Tool Recovery. As Final IPA Wipe Nears Completion, Use More and More Fresh MiraWIPES® to Reduce Chances of Introducing Particles Back Into the Chamber | ||
Figure below shows how much more deposition the Foamtec International MiraWIPE® can remove from a critical surface compared to the standard fab wiper, making the MiraWIPE® FINAL WIPE PROCEDURE the most CRITICAL STEP of the PM procedure (See Fig 22a & 22b) |
||
![]() |
||
Fig 22a: Current fab wiper after completely wiping the LAM 9600 |
Fig 22b: Particles picked up using HT5790S MiraWIPES® after completely wiping with current fab wiper |
|
![]() |
![]() |
|
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY |
||
| Step 37: | With fresh MiraWIPES® and IPA, thoroughly wipe down all areas throughout removable parts |
| Step 38: | Using a single HT5790S MiraWIPE®, folded into quarters, moisten with IPA (See Fig. 23) |
![]() |
||||||
Fig 23: Use MiraWIPE® with IPA for final tool wipe down |
||||||
| Step 39: | Working from highest point down; proceed to effectively wipe out entire Metal ETCH chamber, refolding MiraWIPE® to expose fresh side of MiraWIPE® | |||
| Step 40: | Continue to repeat this sequence until entire Metal ETCH chamber is effectively wiped out | |||
| TOOL RECOVERY: | ||||
| Step 41: | Ensure all PM procedures have been completed and checked off to the point of preparation for tool recovery | |||
| Step 42: | Just prior to closing chamber lid, effectively wipe out entire chamber one last time using a fresh MiraWIPE® with IPA (See Fig 24) | |||
![]() |
||||
Fig 24: Wipe entire chamber with IPA & MiraWIPES® just prior to closing chamber lid |
||||
| Step 43: | Close and seal chamber lid | ||