| OBJECTIVE: | ||
TO EFFECTIVELY PM THE NOVELLUS INOVA COOL STATION CHAMBER IN A TIMELY MANNER, WHILE IMPROVING TOOL RECOVERY AND PARTICLE PERFORMANCE |
||
| Vacuum Chamber: | NOVELLUS INOVA COOL STATION | |
| Vacuum Chamber Process Residue: | PROCESS INDUCED RESIDUE | |
| Vacuum Chamber Components: | CHAMBER |
| Old Procedure: | 1 - 2 hours using DI water & IPA with 30+ wipes (Occasionally use Scotch-Brite™) Tool Recovery: Extended out gassing and conditioning wafers to pass particle specs |
|
| New Procedure: | <1 hour using 360D ScrubPAD, UltraSOLV® Sponge, DI water & IPA w/ MiraWIPES® & MiraSWABS® Tool Recovery: IMPROVED TOOL RECOVERY WITH LESS OUT GASSING AND CONDITIONING WAFERS TO PASS PARTICLES |
|
![]() |
|||
| Vacuum Chamber Products: | |||
NOVELLUS PVD CHAMBER PM KIT |
|||
![]() |
|||
| Novellus Inova Cool Station Chamber PM Procedure: | ||||||||||||||
| View “How to” instructional videos on http://foamtecintlwcc.com/video/ | ||||||||||||||
| Step 1: | Using proper procedures and safety guidelines, shutdown and prepare Novellus INOVA COOL Station chamber for wet clean | |||||||||||||
| Step 2: | Take a certified fab vacuum and vacuum out any loose flaking throughout the cool station chamber | |||||||||||||
| SCRUBBING TOP OF COOL STATION CHAMBER: | ||||||||||||||
| NOTE: | IF IT IS NOT NECESSARY TO SCRUB THE TOP EDGE OF THE COOL STATION CHAMBER, THEN PROCEED TO STEP 11 FOR COOL STATION CHAMBER WIPE PROCEDURE (See Fig 1) | |||||||||||||
|
||||||||||||||
| Step 3: | Take the HT4536D 360 Grit Diamond ScrubPAD and HT4754 UltraSOLV® Sponge and dampen with DI water (See Fig 2) | ||||||||||
|
|||||||||||
| Step 4: | Take lightly dampened UltraSOLV® Sponge and wipe the top edge of cool station chamber that is to be scrubbed (See Fig 3) | ||||||||||||||||||
|
|||||||||||||||||||
| Step 5: | Take lightly dampened 360 Grit Diamond ScrubPAD and scrub buildup along the top edge of the cool station chamber (See Fig 4) | ||||||||||||||||||
|
|||||||||||||||||||
| NOTE: | FOR BEST RESULTS, SCRUB IN ONE DIRECTION ALONG THE OUTER EDGE OF THE O-RING GROOVE. REMOVE AND REPLACE O-RING IF NEW ONE IS AVAILABLE | ||||||||||||||||||
| NOTE: | REMEMBER IT IS NOT NECESSARY TO USE A LOT OF DI WATER DURING THIS SCRUB PORTION OF THE PM, ONLY ENOUGH TO KEEP DIAMOND SCRUBPAD MOIST | ||||||||||||||||||
| Step 6: | After scrubbing a small area with the 360 Grit Diamond ScrubPAD, take the lightly dampened UltraSOLV® Sponge and wipe build up from the scrubbed area | ||||||||||||||||||
| NOTE: | THIS STEP WILL ELIMINATE USING EXCESS WIPERS AS THE SPONGE CAN BE REUSED DURING THE SCRUB PORTION OF THE PM | ||||||||||||||||||
| Step 7: | As Diamond ScrubPAD loads-up with deposition, pull ScrubPAD across damp UltraSOLV® Sponge to properly unload (See Fig 5, 6 & 7) | ||||||||||||||||||
|
|||||||||||||||||||
| Step 8: | Unload the UltraSOLV® Sponge by rinsing out thoroughly with DI water (See Fig 8 & 9) | ||||||||||||||||||
|
|||||||||||||||||||
| Step 9: | Repeat steps 3 – 8 for polishing the entire top edge of the cool station chamber. Unload ScrubPAD and UltraSOLV® Sponge as necessary | ||||||||||||||||||
| Step 10: | When finished with scrub on top edge of cool station chamber, rinse out UltraSOLV® Sponge with fresh DI water and wipe out the entire chamber using the dampened UltraSOLV® Sponge | ||||||||||||||||||
| NOTE: | IF IT WAS NECESSARY TO SCRUB ANY AREA WITHIN THE COOL STATION CHAMBER, REMOVE AND REPLACE YOUR GLOVES WITH A NEW PAIR OF GLOVES PRIOR TO WIPING THE COOL STATION CHAMBER | ||||||||||||||||||
| NOTE: | IF IT WAS NECESSARY TO SCRUB ANY AREA WITHIN THE COOL STATION CHAMBER, TAKE CLEAN N2 OR He TO BLOW OUT THE SMALL AREAS AND TIGHT CORNERS WHERE DI WATER MAY HAVE BEEN CAPTURED | ||||||||||||||||||
IMPORTANT NOTE |
|||||||
| MUST USE HT5790S MiraWIPES® AND HT1511FC MiraSWABS® DURING WIPE PORTION OF PROCEDURE TO EFFECTIVELY REMOVE PARTICLE DEFECTS FROM NOVELLUS INOVA COOL STATION CHAMBER | |||||||
| NOTE: | Below is an example of the particles the HT1511FC MiraSWAB® was able to remove from the hard to reach tight areas in the Novellus INOVA COOL STATION Chamber (See Fig 10 & 11) | ||||||
|
|||||||
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY |
||
| WIPING COOL STATION CHAMBER | ||
| NOTE: | Important to use HT1511FC MiraSWABS® before HT5790S MiraWIPES® as the MiraSWABS® will be lifting particles out of the hard to reach areas and will have to follow wiping with the MiraWIPE® | |||||||||
| Step 11: | Dampen the HT1511FC MiraSWAB® with IPA and wipe out the hard to reach areas, such as o-ring grooves, screw holes and tight corners (See Fig 12 & 13) | |||||||||
|
||||||||||
Figure below shows how much more deposition the Foamtec International MiraWIPE® can remove from a critical surface compared to the standard fab wiper, making the MiraWIPE® FINAL WIPE PROCEDURE the most CRITICAL STEP of the PM procedure (See Fig 24a & 24b) |
||
![]() |
||
Fig 24a: Current fab wiper after completely wiping chamber |
Fig 24b: Particles picked up using HT5790S MiraWIPES® after completely wiping with current fab wiper |
|
![]() |
![]() |
|
MiraWIPES® are the KEY STEP for DEFECT REDUCTION and IMPROVED TOOL RECOVERY |
||
| Step 12: | Fold the HT5790S MiraWIPE® into quarters and dampen with IPA | |||||||||||||
| NOTE: | IMPORTANT TO FULLY SATURATE THE MiraWIPE® WITH IPA; THE MICROFIBER CHARACTERISTICS ARE MORE ABSORBENT AND WILL HOLD MORE IPA THAN ANY STANDARD FAB WIPER HELPING TO PERFORM A MORE EFFECTIVE FINAL WIPE | |||||||||||||
| Step 13: | With the IPA dampened MiraWIPE®, wipe down all areas of the cool station chamber, ensuring to refold the MiraWIPE® as necessary to expose a clean side of the MiraWIPE® as you wipe within the chamber | |||||||||||||
| NOTE: | IT IS GOOD PRACTICE TO WIPE A CHAMBER FROM THE TOP DOWN AND FRONT TO BACK | |||||||||||||
| NOTE: | REPLACE WITH A NEW DAMPENED MiraWIPE® AS NECESSARY | |||||||||||||
| Step 14: | Repeat above MiraWIPE® & MiraSWAB® FINAL WIPE PROCEDURE on all remaining areas of Novellus INOVA Cool Station chamber | |||||||||||||
| NOTE: | IT IS EXTREMELY CRITICAL TO PERFORM THE FINAL WIPE ON THE COOL STATION CHAMBER PLATEN WITH THE UltraSOLV® FOAM WIPER TO PREVENT GENERATING AND LEAVING FIBERS IN THE COOL STATION CHAMBER (SEE FIG 15) | |||||||||||||
|
||||||||||||||
| Step 15: | Take the UltraSOLV® foam wiper and fold into quarters (See Fig 16 & 17) | |||||||||||||
|
||||||||||||||
| Step 16: | Dampen the UltraSOLV® foam wiper with IPA and wipe the cool station chamber platen by PULLING the foam wiper across the platen | |||||||||||||
| NOTE: | IT IS NECESSARY TO PULL THE FOAM WIPER ACROSS THE PLATEN AS THE CHARACTERISTICS OF THE FOAM WIPER WILL NOT ALLOW THE WIPER TO BE PUSHED | |||||||||||||
| NOTE: | REFOLD FOAM WIPER AS NECESSARY TO EXPOSE A NEW AREA OF THE WIPER AFTER EACH SECTION HAS BEEN USED | |||||||||||||
| Step 17: | Using proper procedures and safety guidelines bring cool station chamber back up to production | |||||||||||||