| OBJECTIVE: |
| TO EFFECTIVELY REMOVE POLYMER ON ESC AND OXIDE RESIDUES ON CHAMBER WALL IN A TIMELY MANNER WHILE MINIMIZING PARTICLE ISSUES AND IMPROVING TOOL PERFORMANCE |
| |
| |
Vacuum Chamber: |
TEL DRM ETCH TOOL WITH CERAMIC, ANODIZED ALUMINUM AND POLYAMIDE SURFACES |
| Vacuum Chamber Process Residue: |
PROCESS INDUCED RESIDUE |
| Vacuum Chamber Components: |
DRM CHAMBER, EXIT AND ENTRY PORTS, COMPONENTS |
| |
| |
Old Procedure: |
Standard cleanroom polyester wipers and Scotch-Brite™ scrub pads |
|
Solvent: |
DI water, IPA |
| |
|
|
|
| |
Vacuum Chamber Products: |
|
|
| |
| TEL Unity 2 DRM Chamber PM Procedure: |
|
| View “How to” instructional videos on http://www.foamtecintlwcc.com/video/ |
| |
| Step 1: |
After the chamber has been vacuumed, clean the chamber with a lightly dampened HT4754 UltraSOLV® Sponge in order to remove any loosely adhered residues |
| |
| Step 2: |
Cover the ESC to prevent scratching. Lightly scrub the upper ceramic areas with the HT4540A-10 UltraSOLV® ScrubPAD to remove residue. Occasionally wipe scrubbed areas with HT4754 UltraSOLV® Sponge to remove loosened residue. The same sponge can also be used to “unload” the ScrubPAD for maximum effectiveness (See Fig 1, 2 & 3) |
| |
|
Fig 1: ScrubPAD loaded with deposition |
|
|
Fig 2: Pull ScrubPAD across UltraSOLV® Sponge |
|
|
Fig 3: Unloaded ScrubPAD |
|
|
| Step 3: |
Unload UltraSOLV® as much as possible by placing it in container of DI water and RINSE-OUT thoroughly (See Fig 4 & 5) |
| |
| |
Fig 4: UltraSOLV® Sponge loaded with deposition |
|
|
Fig 5: UltraSOLV® Sponge free of deposition after rinse in DI water |
|
|
| |
| Step 4: |
Move the pedestal to process position to access areas below and around the ESC. Using a fresh UltraSOLV® ScrubPAD, lightly scrub the vertical walls of the pedestal to remove excess residue. As build-up is loosened, follow with Sponge wipe for maximum effectiveness. Remember to unload the UltraSOLV® ScrubPAD frequently to increase cleaning efficiency. Repeat above for remaining pedestal areas |
| |
| NOTE: |
FOR PEDESTAL WALLS WITH EXTREMELY HARD RESIDUE, IT IS RECOMMENDED TO USE AN HT4580D-10, 800 GRIT DIAMOND ULTRASOLV® SCRUBPAD. FOR HARD-TO-REACH AREAS AND SMALL VERTICAL SURFACES USE AN HT174813PD SCRUBTIP®. AN HT179060 SCRUBTIP® SHOULD BE USED TO CLEAN VERTICAL SURFACES. TO FINAL CLEAN THESE AREAS USE ENOUGH HT1748 AND HT1790 ULTRASOLV® SWABS UNTIL NO VISIBLE CONTAMINATION TRANSFERS OFF SURFACE TO THE SWAB |
| |
| Step 5: |
Use an HT173280 ScrubTIP® to clean areas where the bafflers are fastened to the chamber. HT1732 UltraSOLV® Swabs should be used where residue has been deposited |
| |
| Step 6: |
Vacuum the scrubbed areas to ensure final removal of loosened particles |
| |
| Step 7: |
Finally, wipe down chamber with 25-40 HT4790 UltraSOLV® Wipers and isopropyl alcohol |
| |
| NOTE: |
IF THE BAFFLE AND CERAMIC RINGS NEED TO BE CLEANED IN PLACE, THE FOLLOWING PROCEDURE SHOULD BE USED |
| |
| |
Step: A |
The ceramic focus ring can be cleaned with an HT4536D-10 UltraSOLV® Diamond ScrubPAD. To final clean ceramic, wipe with HT4790 UltraSOLV® Wipers until all residue is removed |
| |
| |
Step: B |
The baffle is easily cleaned with HT174813PD ScrubTIP®s. The ScrubTIPs® are also effective at cleaning loose residues from the tight spaces in the “fan sector” area. Remember to clean both sides of the baffle with a fresh UltraSOLV® ScrubPAD |
Scotch-Brite™ is a trademark of 3M Corporation